Remote Control Commands
R&S
®
SMBV100A
668
Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:LIST:SELect
<Select>
The command selects the specified list. If a new list is to be created, the name can be
entered here. The list is created if it does not yet exist. The list selected here is available
for the further processing steps (editing) and is used in the instrument when the List mode
is activated.
The files are stored with the fixed file extensions *.lsw in a directory of the user's choice.
The directory applicable to the command is defined with the command
MMEMory:CDIR
.
A path can also be specified in command
:SOUR:LIST:SEL
in which case the list mode
file in the specified directory is selected.
*RST does not affect data lists.
Parameters:
<Select>
'<list name>'
Example:
LIST:SEL '/var/smbv/Lists/Listmode/list3'
selects list3 for editing. List3 is created if it does not yet exist.
[:SOURce<hw>]:LIST:TRIGger:EXECute
The command immediately starts the processing of a list in list mode. It corresponds to
the manual-control command "Execute Single."
Example:
SOUR:LIST:TRIG:EXEC
triggers the processing of the selected list.
Usage:
Event
[:SOURce<hw>]:LIST:TRIGger:SOURce
<Source>
The command sets the trigger source for the LIST mode.
The names of the parameters correspond to those under sweep mode. SCPI uses other
names for the parameters; these names are also accepted by the instrument. The SCPI
names should be used if compatibility is an important consideration. An overview of the
various names is given in the following table:
R&S name
SCPI name
Command under manual control
AUTO
IMMediate
MODE AUTO
SINGle
BUS
MODE SINGLE
or
STEP
EXTernal
EXTernal
MODE EXT TRIG SINGLE
or
EXT
TRIG STEP
SOURce Subsystem