Clock Settings
This section lists the remote control commands, necessary to configure the clock.
The clock settings are available for R&S SMx and R&S AMU instruments only.
......................................................................608
................................................................608
...................................................................609
............................................609
.........................................610
[:SOURce<hw>]:BB:MCCW:CLOCk:MODE
<Mode>
The command enters the type of externally supplied clock (
:BB:MCCW:CLOCk:SOURce
EXTernal
). When
MSAMple
is used, a multiple of the sample clock is supplied via the
CLOCK connector and the sample clock is derived internally from this. The multiplier is
entered with the command
BB:MCCW:CLOCk:MULTiplier
.
Parameters:
<Mode>
SAMPle | MSAMple
*RST:
SAMPle
Example:
BB:MCCW:CLOC:MODE SAMP
selects clock type "Sample", i.e. the supplied clock is a sample
clock.
[:SOURce<hw>]:BB:MCCW:CLOCk:MULTiplier
<Multiplier>
The command specifies the multiplier for clock type "Multiple Samples"
(
:BB:MCCW:CLOCk:MODE MSAM
) in the case of an external clock source.
Parameters:
<Multiplier>
integer
Range:
1 to 64
Increment: 1
*RST:
4
Example:
BB:MCCW:CLOC:SOUR EXT
selects the external clock source. The clock is supplied via the
CLOCK connector.
BB:MCCW:CLOC:MODE MSAM
selects clock type "Multiple Samples", i.e. the supplied clock has
a rate which is a multiple of the sample rate.
BB:MCCW:CLOC:MULT 12
the multiplier for the external clock rate is 12.
SOURce Subsystem