Instrument Function
R&S
®
SMBV100A
146
Operating Manual 1407.6062.32 ─ 08
SCPI command:
Mode
Selects the graphical signal display. The display is not shown until it has been activated
with "Status On".
The individual signal displays are described in the
SCPI command:
Trigger Source
Defines the trigger for the starting time of the graphic recording.
"Software"
Recording of signals is started automatically in specified intervals.
This asynchronous method is appropriate when a representative dis-
play of the complete signal is desired, as recording starts in a random
time reference to the signal.
"Marker 1"
The starting point for recording signals is determined by marker 1 in the
menu of the active standard.
This synchronous method is appropriate when specific signal segments
are to be shown in greater detail. With periodic signals, a static image
is obtained by selecting a period length of marker 1 that is equal to the
signal period.
The displayed signal segment can be shifted as needed by entering a
marker delay in the Trigger/Marker menu of the active standard.
SCPI command:
Scrambling Code
(only if Code Domain (3GPP FDD DL) is selected)
Sets the scrambling code if the "Code Domain" display is selected. Since it is possible to
select a different scrambling code in the menu for each of the 4 base stations and it is
also possible to display an external signal, the scrambling code for which the display is
to be generated must be explicitly specified here.
SCPI command:
n.a.
Sample Rate Mode
Sets how the time resolution of the signal is determined. Maximum resolution corre-
sponds to a display covering the entire signal bandwidth. The higher the resolution is, the
shorter the length of the displayed signal segment will be for the specified recording
depth.
Graphical Display - Graphics