Instrument Function
R&S
®
SMBV100A
349
Operating Manual 1407.6062.32 ─ 08
If the combined waveforms have different sample rates, they can be adapted to a com-
mon sample rate by resampling. By the same procedure the instantaneous amplitude of
the various waveforms can be scaled to a common rms level.
The configuration of a multi segment waveform, that is to say details of how it is made
up from different waveforms, the level and clock rate settings, and the file name, can be
saved separately in a list. The file extension is
*.inf_mswv
. This method can be used
to create any number of configurations as a basis for defining further multi segment
waveforms.
Example for the creation of a multi segment waveform file.
1. Select "New List" to create empty list.
2. Use the "Append" function to add two or more waveform files.
3. Adjust the "Level", "Clock" and "Segment Marker" settings.
4. Select the "Output File" and enter file name.
5. Select "Save List" to save the configuration.
6. Select "Create" or "Create and Load" to create and save the multi segment waveform.
Multi Segment File Contents
The upper part of the dialog provides the functions, necessary to create and adjust the
contents of the multi segment file.
Baseband Signal - Baseband Block