Remote Control Commands
R&S
®
SMBV100A
596
Operating Manual 1407.6062.32 ─ 08
Return values:
<Clock>
float
Range:
0 to Max
Increment: 1E-3
*RST:
0
Example:
BB:MCCW:CLOC?
queries the output clock rate.
Response:
256 000 000
the output clock rate is 256
MHz.
Usage:
Query only
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:EXECute
The command adopts the settings for the carrier range which has been defined using
the
:BB:MCCW:EDIT:CARR:...
commands.
Example:
BB:MCCW:EDIT:CARR:STAR 4
the carrier range starts at carrier 4.
BB:MCCW:EDIT:CARR:STOP 400
the carrier range stops at carrier 400.
BB:MCCW:EDIT:CARR:STAT ON
sets all the carriers in the carrier range to ON.
BB:MCCW:EDIT:CARR:EXEC
adopts the settings for all the carriers in the carrier range.
BB:MCCW:STAT
starts generation of the multi carrier signal. Carriers 4 to 400 are
in the ON state.
Usage:
Event
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe:STEP
<Step>
The command sets the step width by which the start phases of the carriers in the defined
carrier range will be incremented.
The phase settings are only valid if optimization of the crest factor is disabled
(
:SOURce:BB:MCCW:CFACtor:MODE OFF
).
Parameters:
<Step>
float
Range:
-359.99 to 359.99
Increment: 0.01
*RST:
0
Example:
BB:MCCW:EDIT:CARR:PHAS 90 DEG
sets a start phase of 90° for the carriers in the carrier range.
BB:MCCW:EDIT:CARR:PHAS:STEP 1 DEG
the start phase is incremented by 1° for each carrier, i.e. the first
carrier has a start phase of 90°, the second a start phase of 91°,
etc.