Instrument Function
R&S
®
SMBV100A
279
Operating Manual 1407.6062.32 ─ 08
To easy the configuration, an automatically adjustment of the required synchronization
settings is provided.
Step 1: Adjust Phase and Level
► Refer to the Application Note 1GP67 for a detailed description on how to adjust the
phase and level settings manually or by using the software tool PhaseTracker.
Step 2: Configure Master Instrument
For working in master-slave synchronous mode, configure the master instrument as fol-
low:
1. Adjust the baseband settings (e.g. "Custom Digital Modulation" or "ARB") as required
and activate signal generation.
2. Set the "Synch Mode" to Sync Master.
SCPI Command:
BB:DM:CLOC:SYNC:MODE MAST
3. Perform "Set Synchronization Settings".
SCPI Command:
BB:DM:CLOC:SYNC:EXEC
The synchronization settings are automatically set.
The synchronization clock (including the trigger signal) and the reference clock of the
master instrument are output on the CLK OUT and the REF OUT connectors and
supplied to the slave instrument.
Step 3: Configure Slave Instrument(s)
For working in master-slave synchronous mode, the slave instrument(s) has (have) to be
configured as follow:
1. Configure the instrument to generate a baseband signal with the same settings as
the master one and activate signal generation.
2. Adjust the state of the parameter "Trigger/Marker/Clock > Sync. Output to External
Trigger" as follow:
a) Disable the parameter, if the master instrument provides the slaves with its inter-
nal trigger signal.
b) In case a common external trigger signal is provided for the master and the slave
instrument(s), the state of this parameter in all instruments has to match.
SCPI Command:
BB:DM:TRIG:EXT:SYNC:OUTP OFF
3. Set the "Sync Mode" to "Sync Slave".
Baseband Signal - Baseband Block