Instrument Function
R&S
®
SMBV100A
145
Operating Manual 1407.6062.32 ─ 08
5.3.1 Graphics Settings Menu
The "Graphics Settings" menu for selecting the graphical display of the output signal is
opened either in the "Graphics" function block or in the menu with the same name which
is opened using the MENU key.
The signal display can be selected and activated in the menu.
With two-path instruments, the settings for the two paths can be entered separately. The
signals can be displayed simultaneously in two separate windows.
State
Activates the selected graphical display.
After activation, the diagram selected with "Mode" is displayed in the block diagram.
At the bottom of each graphics window there are several buttons for freezing and zooming
the display and for activating a reference curve.
For more information, see
chapter 5.3.2, "Bar of Graphics Window"
SCPI command:
Smart Graphics
Activates the "smart graphics".
The graphic is displayed in a small window which is roughly the size of a block. These
windows are displayed permanently in the block diagram until they are deactivated, and
allow basic checking of the signal. They do not contain any buttons.
Graphical Display - Graphics
Summary of Contents for SMBV100A
Page 31: ......