CALIBRATION PROCEDURE
NI PXIe-5601
This document contains information for calibrating the National Instruments PXIe-5601 (NI 5601)
RF downconverter. For more information about calibration, visit
ni.com/calibration
.
Contents
Software Requirements............................................................................................................. 2
Documentation Requirements .................................................................................................. 2
Calibration Interval................................................................................................................... 2
Test Equipment......................................................................................................................... 2
Test Conditions......................................................................................................................... 4
Calibration Procedures ............................................................................................................. 5
Power Splitter Reference Output.............................................................................. 6
Power Sensor Zero/Calibration ................................................................................ 6
RF Input Calibration................................................................................................. 6
IF Output Calibration................................................................................................ 9
Absolute Gain Accuracy Verification ...................................................................... 10
Average Noise Level Verification............................................................................ 17
LO Output Power Verification ................................................................................. 19
Instantaneous Bandwidth Verification ..................................................................... 21
Gain Compression Verification ................................................................................ 24
120 MHz.................................. 28
IF Attenuation Calibration Adjustment .................................................................... 33
IF Response Adjustment........................................................................................... 36
Reference Level Adjustment for RF Frequencies
120 MHz.................................. 41
Appendix A: Reference Level Calibration Power.................................................................... 44
Where to Go for Support .......................................................................................................... 45