![National Instruments NI PXIe-5601 Calibration Procedure Download Page 34](http://html1.mh-extra.com/html/national-instruments/ni-pxie-5601/ni-pxie-5601_calibration-procedure_3444485034.webp)
34
|
ni.com
|
NI PXIe-5601 Calibration Procedure
9.
Lock the NI PXI/PXIe-5652 to the Reference Clock used by the external test equipment.
Use the niRFSG Configure Ref Clock VI or the
niRFSG_ConfigureRefClock
function.
10. Set the NI PXI/PXIe-5652 power level to 0 dBm using the niRFSG Power Level property
or the
NIRFSG_ATTR_POWER_LEVEL
attribute.
11. Set the NI PXI/PXIe-5652 frequency to 187.5 MHz using the niRFSG Frequency property
or the
NIRFSG_ATTR_FREQUENCY
attribute.
12. Commit the settings to hardware using the niRFSG Initiate VI or the
niRFSG_Initiate
function.
13. Call the niRFSA Initiate VI or the
niRFSA_Initiate
function.
14. Tune the RF source to the RF Frequency in Table 21.
15. Configure the NI 5601 RF signal path by setting the niRFSA RF Path Selection property or
the
NIRFSA_ATTR_CAL_RF_PATH_SELECTION
attribute according to the value
referenced in Table 21 to the corresponding values:
•
RF band 2
•
RF band 3
16. Configure the IF filter by setting the niRFSA IF Filter property or
the
NIRFSA_ATTR_IF_FILTER
attribute according to the value referenced in Table 21 to
the corresponding values:
•
53 MHz
•
187.5 MHz, narrow
•
187.5 MHz, wide
17. Configure the NI 5601 center frequency by setting the niRFSA Center Frequency property
or the
NIRFSA_ATTR_SPECTRUM_CENTER_FREQUENCY
attribute using
the
RF Frequency
shown in Table 21.
18. Set the RF source power to -32 dBm.
19. Configure the NI PXI/PXIe-5652 frequency to the
LO Frequency
shown in Table 21 using
the niRFSG Frequency property or the
NIRFSG_ATTR_FREQUENCY
attribute.
20. Wait 750 ms or until power sensor A settles to 0.1% of source power, whichever is greater.
21. Configure the IF attenuation for the device by setting the niRFSA IF1 Attenuation and
IF2 Attenuation properties or the
NIRFSA_ATTR_IF1_ATTEN_VALUE
and
NIRFSA_ATTR_IF2_ATTEN_VALUE
attributes. This configuration ensures that Table 22
references all possible value pairs for IF1 and IF2.
Table 21.
NI 5601 Center Frequency
IF Filter
RF Frequency
LO Frequency
RF Band
53 MHz
200 MHz
253 MHz
3
187.5 MHz, Narrow
1 GHz
1,187.5 MHz
2
187.5 MHz, Wide
1 GHz
1,187.5 MHz
2