NI PXIe-5601 Calibration Procedure
15. Configure the IF filter by setting the niRFSA IF Filter property or
the
NIRFSA_ATTR_IF_FILTER
attribute
to the IF bandwidth value referenced in
Table 23 or 24, corresponding to the module revision of your NI 5601, for each of the
following filters:
•
187.5 MHz, wide
•
187.5 MHz, narrow
•
53 MHz
16. Configure the IF attenuation for the device by setting the niRFSA IF1 Attenuation and
IF2 Attenuation properties or the
NIRFSA_ATTR_IF1_ATTEN_VALUE
and the
NIRFSA_ATTR_IF1_ATTEN_VALUE
attributes. Use the values referenced from Table 23
or 24, corresponding to the module revision of your NI 5601.
17. Configure the NI 5601 center frequency by setting the niRFSA Center Frequency property
or the
NIRFSA_ATTR_SPECTRUM_CENTER_FREQUENCY
attribute. Use the RF frequency
shown in Table 23 or 24, corresponding to the module revision of your NI 5601.
18. Commit the settings to hardware using the niRFSA Initiate VI or the
niRFSA_Initiate
function.
19. Configure the spectrum analyzer using the settings shown in Table 25.
20. Set the RF source frequency to the
RF Frequency
in Table 23 or 24, corresponding to the
module revision of your NI 5601.
Table 24.
IF Response Settings (NI 5601 Module Revisions G and Later)
IF
Filter
RF
Freq.
(MHz)
NI 5652
LO Start
Freq.
(MHz)
NI 5652
LO Stop
Freq.
(MHz)
NI 5652
LO Freq.
Step
(MHz)
NI 5601
LO Freq.
(MHz)
IF1
Atten.
(dB)
IF2
Atten.
(dB)
RF
Band
187.5
MHz,
Wide
1,000
1,307.5
1,067.5
1
187.5
8
3
2
187.5
MHz,
Narrow
1,000
1,227.5
1,147.5
1
187.5
8
3
2
53
MHz
200
305
201
0.5
53
8
3
3
Table 25.
Spectrum Analyzer Settings
Parameter
Value
Center frequency
NI 5601 LO frequency
-
NI 5601 center frequency
Span
10 Hz
Reference level
20 dBm
Resolution bandwidth
200 Hz