NI PXIe-5601 Calibration Procedure
|
© National Instruments
|
17
Average Noise Level Verification
1.
Connect a 50
Ω
terminator to the NI 5601 RF IN connector. Ensure that the 50
Ω
terminator
is connected to the NI 5601 LO OUT connector.
2.
Connect the NI 5601 IF OUT connector to the spectrum analyzer input through the
3.5 mm-to-3.5 mm cable. Figure 5 shows the completed equipment setup.
Figure 5.
Average Noise Level Equipment Setup
Note
The figure shows the NI PXI-5652, but you can also use the NI PXIe-5652.
3.
Create a new session for the NI 5601 using the niRFSA Initialize With Options VI or the
niRFSA_initWithOptions
function. Set the Driver Setup portion of the
option string
parameter to
DriverSetup=Digitizer:<external>
.
4.
Lock the NI PXI/PXIe-5652 to the Reference Clock used by the external test equipment.
Use the niRFSA Configure Ref Clock VI or the
niRFSA_ConfigureRefClock
function.