Network operation and remote control
R&S
®
LCX Series
148
User Manual 1179.2260.02 ─ 02
To read the register, use the query commands
mands for the operation status register are described in
Chapter 11.14.1, "Status oper-
Table 10-3: Assignment of the bits used in the operation status register
Bit No.
Meaning
0
CALibrating
The bit is set during the calibration phase (for service department only).
1
SETting
2
RANging
3
UPDate
4
0
5
TRIGger
This bit is set while the instrument is waiting for the trigger.
6
0
7
0
8
LIMit
9
INTegrating
10
LOGging
This bit is set as long as "Logging" is enabled
11
LOCKed
Instrument locked (RWLock)
12
INRush
13
0
14
0
15
This bit is always 0
10.2.4.3
Instrument-specific status questionable register
This status register contains information on questionable instrument states. Questiona-
ble states occur when the instrument is not operated in compliance with its specifica-
tions.
To read the register, use the commands
STATus:QUEStionable:CONDition?
on page 211. The remote
commands for the status questionable register are described in
Remote control interfaces and protocols