![Rohde & Schwarz R&S LCX Series User Manual Download Page 197](http://html.mh-extra.com/html/rohde-and-schwarz/rands-lcx-series/rands-lcx-series_user-manual_1477867197.webp)
Remote control commands
R&S
®
LCX Series
197
User Manual 1179.2260.02 ─ 02
FUNCtion:IMPedance:SOURce
<Impedance>
FUNCtion:IMPedance:SOURce?
<Impedance>
Selects the output impedance for the measurement.
Parameters for setting and query:
<Impedance>
LOW | R10 | HIGH | R100
LOW | R10
Sets 10
Ω
output impedance.
HIGH | R100
Sets 100
Ω
output impedance.
*RST:
HIGH
Example:
See
Example "Selecting the measurement function and setting
Manual operation:
See
FUNCtion:IMPedance[:TYPE]
<Impedance Type>
FUNCtion:IMPedance[:TYPE]?
<Impedance Type>
Selects the impedance parameter for the measurement corresponding to the measure-
ment type, see
Parameters for setting and query:
<Impedance Type>
CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS | LPD | LPQ |
LPG | LPRP | LSD | LSQ | LSRS | RX | RPB | RDC | MTD |
NTD | ZTD | ZTR | GB | YTD | YTR
CPD | CPQ | CPG | CPRP | CSD | CSQ | CSRS
Capacitive measurement type: Cp (parallel capacitance), Cs
(serial capacitance), D (dissipation factor), Q (quality factor), G
(conductance), Rp (parallel resistance), Rs (serial resistance)
LPD | LPQ | LPG | LPRP | LPRDc | LSD | LSQ | LSRS |
LSRDc
Inductive measurement type: Lp (parallel inductance), Ls (serial
inductance), D (dissipation factor), Q (quality factor), G (conduc-
tance), Rp (parallel resistance), Rs (serial resistance), RDc
(direct current resistance)
RX | RPB | RDC | MTD | NTD | ZTD | ZTR | GB | YTD | YTR
Resistance measurement type: R (resistance), X impedance, Rp
(parallel resistance), RDC (direct current resistance), B (suscep-
tance), M (mutual inductance), N (transformer ratio), Z (impe-
dance), G (conductance), Y (admittance), TD (phase angle
degree), TR (phase angle rad)
Example:
See
Example "Selecting the measurement function and setting
Manual operation:
See
FUNCtion subsystem
www.allice.de
Allice Messtechnik GmbH