Getting started
R&S
®
LCX Series
31
User Manual 1179.2260.02 ─ 02
Power supply connector for connecting the R&S
LCX to the mains., see
ter 4.1.6, "Connecting to power"
Ground terminal
Protective ground terminal to secure the R&S
LCX, e.g. with a grounded external con-
ductor, see
Digital I/O
15-pole D-Sub socket to connect the binning interface.
External Voltage Bias
4
mm safety sockets for external bias voltage.
Note:
External BIAS requires constant voltage.
Before applying the voltage, make sure that you have set the constant voltage mode in
the supplying instrument. Refer to the user manual of the power supply.
The External Voltage Bias socket is not directly connected to ground.
Indirectly, there is a connection to ground with low impedance over the "H CUR" output
buffer.
When the bias voltage is turned on, the potential to ground deviates only by a maxi-
mum of
±
18
V. Without external bias, make sure that the voltage at the two sockets
does not exceed the safety extra low voltage or the specified 40
V
DC.
BIAS Fuse
Socket for the fuse securing the external bias voltage.
Trigger Input
BNC socket for external trigger signal.
The trigger input is directly connected to the mains safety ground by the line cord.
USB A
Female USB (universal serial bus) connector of type A (host USB). You can connect a
USB memory stick, e.g., to record and export measurement data, to capture screen-
shots, or to update the firmware.
How to:
Chapter 4.1, "Preparing for use"
USB B
Female USB (universal serial bus) type B connector. This interface provides remote
control of the instrument from a controller PC.
"Establishing a USB remote control connection"
LAN
RJ-45 socket to connect the R&S
LCX to a LAN for remote control, remote operation,
and data transfer.
How to:
Chapter 4.1.7, "Connecting to LAN"
Instrument tour