Remote control commands
R&S
®
LCX Series
181
User Manual 1179.2260.02 ─ 02
Example: Setting correction prameters
The example represents the remote control commands to configure the correction set-
tings.
// Configure the corrections settings:
// set the cable length between the DUT and the instument
CORRection:LENGth 1 // the test leads are of 1m legth
// ***************************
// Perform open correction on all frequencies
CORRection:OPEN:STATe 1
CORRection:OPEN[:EXECute]
// ***************************
// Specifiy a dedicated working point
// e.g. second point at 1000Hz
// Perform open correction on the defined working point
CORRection:SPOT<Spot>:STATe 2,1
CORRection:SPOT<Spot>:FREQuency 2,1000
CORRection:SPOT<Spot>:OPEN[:EXECute] 2
// ***************************
// Perform short correction on all frequencies,
// and at the working point, no 2 1000Hz
CORRection:SHORt:STATe 1
CORRection:SHORt[:EXECute]
CORRection:SPOT<Spot>:FREQuency 2,1000
CORRection:SPOT<Spot>:STATe 2,1
CORRection:SPOT<Spot>:SHORt[:EXECute] 2
// ***************************
// Perform load correction on all frequencies,
// select load type: LPD (parallel inductance, dissipation factor)
CORRection:LOAD:STATe 1
CORRection:LOAD:TYPE LPD
CORRection:SPOT<Spot>:LOAD:STANdard 2,15,20
// Perform load correction at a single working point, no 2 at 1000Hz
CORRection:SPOT<Spot>:FREQuency 2,1000
CORRection:SPOT<Spot>:STATe 2,1
CORRection:SPOT<Spot>:LOAD:STANdard 2,10
CORRection:SPOT<Spot>:LOAD[:EXECute]
// ***************************
// Query the current status of correction
CORRection:OPEN:STATe?
CORRection subsystem