Instrument functions
R&S
®
LCX Series
76
User Manual 1179.2260.02 ─ 02
Serial/Paralell Separation Point
...................................................................... 77
Measurement Speed
Selects the speed of the measurement repetitions per second.
Using this speed parameter requires that you have set continuous triggering, see
Chapter 8.1.1, "Measurement mode settings"
"Fast:
≤
15
ms"
Number of measurements:
≈
15
/
s.
"Medium:
≤
100
ms"
Number of measurements:
≈
10
/
s.
"Slow:
≤
500
ms"
Number of measurements:
≈
1.5
/
s.
Remote command:
Cable Length
Selects the length of the leads to the connected test fixture, i.e. to the DUT.
When connected to a test fixture with leads, the R&S
LCX considers the cable losses.
If you have the test fixture connected to the BNC measurement connectors directly,
select 0
m, as there are no cable losses to be compensated. For test adapters with
leads, set the parameter to 1
m accordingly.
Remote command:
Test Signal Type
Selects the parameter for the setting the level of the test signal level.
The setting parameter for the test signal level on the measurement window varies
according to the selection, see
"Voltage"
Enables you to set the signal level in volt.
"Current"
Enables you to specify the signal current.
Remote command:
Voltage Bias
Activates the use of an external voltage bias, if supplied.
Note:
Make sure that the external DC voltage source supplies a constant voltage bias
to avoid distortion of the measurement results due to interference.
To set the value, see
Chapter 6.3.2, "External voltage bias"
Remote command:
Source Impedance
Selects the output impedance for the measurement.
Measurement parameters