Instrument functions
R&S
®
LCX Series
80
User Manual 1179.2260.02 ─ 02
Remote command:
Start Load Correction (Spot)
Figure 8-6: Start load correction at one working point with specified reference value
Executes the load correction at a defined working point.
The function requires that you have activated the state, see
on page 79 and specified the real resistance and the complex reactance
of the reference component. The R&S
LCX, prompts you to connect the specified load.
"Aligning the instrument using load correction"
Remote command:
CORRection:SPOT<Spot>:LOAD[:EXECute]
Reference Load R
Sets the resistance value as refence for the load correction.
Reference Load X
Sets the impedance value as refence for the load correction.
8.6.3
Setting measurement parameters
The following procedures provide a brief overview of the steps required to perform the
corrections. They point out special features, such as load correction.
Previous steps
We assume that you have prepared the measurement as follows:
1. You have set up and turned on the R&S
Chapter 4.1, "Preparing for use"
2. You have set the R&S
3. You have connected the test fixture, see
Chapter 4.1.9, "Connecting a test fixture"
Tip:
For information on how to handle the connection pins of the individual test fix-
tures during correction, see
Chapter 6.1, "About test fixtures"
Measurement parameters