General instrument settings
R&S
®
LCX Series
130
User Manual 1179.2260.02 ─ 02
"Model"
Designation of the instrument.
"ID"
Part number of the instrument.
"Serial No.:"
Serial number of the instrument.
"Version"
Currently installed firmware version.
How to: see
"FPGA"
Version of the installed FPGA.
"Hardware IDs"
Unique serial number of the front and mainboard of the instrument.
"Temperature in °C"
Note:
If the temperature exceeds the specification, the R&S
LCX dis-
plays a warning message.
Temperature of the CPU, connector board, H
CUR and L
CUR output
connectors, low current supply and Rfb (feedback resistor), indicated
in degrees.
"Misc"
Fan speed, IP address and memory capacity of the instrument.
Remote command:
Create Dump File
Saves the device information in a file.
9.10.2
Using the dump file function
The dump file function allows you to save the device information in the file format
*.rsd
. It saves data in a structured format. You can, export this file and forward it to
the service department, e.g. for analyzing or troubleshooting. Proceed as described in
the following example.
Creating a dump file
To create and export a device information file:
1. Press the [settings] key at the front panel.
2. In the "Device" tab, select "Device Information".
3. Select "Create Dump File".
The R&S
LCX generates a directory in the internal memory and saves the file
under
/int/service/service.rsd
.
4. To export the file:
a) Connect a USB memory device at the front panel USB B connector.
b) Press the [settings] key.
c) Select "File Manager".
d) In the internal memory, select
/int/service/
.
e) Select
/int/service/service.rsd
.
f) Copy the file with the copy left softkey to the USB directory.
Device information
www.allice.de