Remote control commands
R&S
®
LCX Series
172
User Manual 1179.2260.02 ─ 02
11.2
Programming examples
The corresponding sections of the same title provide simple programming examples for
the R&S
LCX. The purpose of the examples is to present
all
commands for a given
task. In real applications, one would rather reduce the examples to an appropriate sub-
set of commands.
The programming examples have been tested with a software tool which provides an
environment for the development and execution of remote tests. To keep the examples
as simple as possible, only the "clean" SCPI syntax elements are reported. Non-exe-
cutable command lines (for example comments) start with two // characters.
At the beginning of the most remote control programs, we recommend that you
(p)reset the instrument to set the R&S
LCX to a definite state. The commands
*RST
and
SYSTem:PRESet
are equivalent for this purpose.
*CLS
also resets the status reg-
isters and clears the output buffer.
In all the examples we assume that:
●
A remote PC is connected to the instrument
●
The remote PC and the instrument are switched on
●
A connection between them is established
11.3
Common commands
Common commands are described in the IEEE 488.2 (IEC 625-2) standard. These
commands have the same effect and are employed in the same way on different devi-
ces. The headers of these commands consist of "*" followed by three letters. Many
common commands are related to the Status Reporting System.
We recommend that you reset the R&S
LCX with
on page 174 to set the instru-
ment to a defined status before starting any remote control sequence.
The R&S
LCX does not support parallel processing of remote commands. If the query
*OPC? (see
on page 173) returns a „1“, the device is able to process new com-
mands.
Available common commands:
Common commands