Remote control commands
R&S
®
NGP800
171
User Manual 5601.5610.02 ─ 09
LOG:DATA?
Returns 12 sets of latest logging data with minimum logging interval (8 ms).
Depending on the models, the data is returned in the following format for a 2-channel
models: <Ch1_voltage>, <Ch1_current>,<Ch1_power>,<Ch2_voltage>, <Ch2_cur-
rent>,<Ch2_power>, <Ch1_voltage>, <Ch1_current>,<Ch1_power>, <Ch2_voltage>,
<Ch2_current>,<Ch2_power>...
Example:
2-channel models with 5 V set on Ch1 and 10V set on Ch2
LOG:DATA?
->
"5.000,0.0000,0.000,10.000,0.0000,0.000,5.000,0.0000,0.000,1
0.000,0.0000,0.000,5.000,0.0000,0.000,10.000,0.0000,0.000,5.
000,0.0000,0.000,10.000,0.0000,0.000,5.000,0.0000,0.000,10.0
00,0.0000,0.000,5.000,0.0000,0.000,10.000,0.0000,0.000,
5.000,0.0000,0.000,10.000,0.0000,0.000,5.000,0.0000,0.000,
10.000,0.0000,0.000,5.000,0.0000,0.000,10.000,0.0000,0.000,
5.000,0.0000,0.000,10.000,0.0000,0.000,5.000,0.0000,0.000,10
.000,0.0000,0.000,5.000,0.0000,0.000,10.000,0.0000,0.000"
Usage:
Query only
LOG:DURation
<Set new value>[, <Return min or max>]
LOG:DURation?
[<Return min or max>]
Sets or queries the duration of the data logging.
Setting parameters:
<Set new value>
<numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Duration of the data logging captured in the range of 0 s to
3.49*10^5 s.
MIN | MINimum
Minimum duration of the data logging captured at 0 s.
MAX | MAXimum
Maximum duration of the data logging captured at 3.49*10^5 s.
Default unit: s
Parameters for setting and query:
<span>
MIN | MINimum | MAX | MAXimum
Returns the duration of the data logging.
Example:
LOG:DUR MAX
LOG:DUR? MAX
-> 349000
LOG:FNAMe
<Set new value>
Sets or queries the filename and storage location for the data logging.
Setting parameters:
<Set new value>
Data and file management commands