Remote control commands
R&S
®
NGP800
165
User Manual 5601.5610.02 ─ 09
State
Descriptions
0-15
0x0 - 0xE ( 0x0000 - 0x1111)
bit3|bit2|bit1|bit0
bit0 - pin 1 of analog input
bit1 - pin 2 of analog input
bit2 - pin 3 of analog input
bit3 - pin 4 of analog input
e.g. 15 - All analog input pins are adjusted.
e.g. 9 - Pin 1 and pin 4 are adjusted.
16
Idle
17
Busy
18
Waiting
Example:
CAL:AINP:STAT?
->
9
Pin 1 and pin 4 are adjusted successful.
Usage:
Query only
CALibration:AINPut:UMAX
Sets output voltage to high value 100 % of Vmax for analog input pin during adjust-
ment.
Usage:
Event
CALibration:AINPut:UMIN
Sets the output voltage to low value 1 % of Vmax for analog input pin during adjust-
ment.
Usage:
Event
CALibration:CANCel
Cancels the channel adjustment.
Usage:
Event
CALibration:COUNt?
Queries the number of counts channel adjustment performed successfully.
Usage:
Query only
Advanced operating commands