Remote control commands
R&S
®
NGP800
122
User Manual 5601.5610.02 ─ 09
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit[:UPPer]
<New value for current>[,
<Channel list>]
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit[:UPPer]?
[<Channel list>]
Sets or queries the upper safety limit for current.
Setting parameters:
<current>
<numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Numeric value for upper safety limit.
MIN | MINimum
Min value for upper safety limit.
MAX | MAXimum
Max value for upper safety limit.
Range:
For up to 32V: 0.0005E+00 to 20.0100E+00. For up
to 64V: 0.0005E+00 to 10.0100E+00
Increment: 0.0005
*RST:
0.0005E+00
Parameters for setting and query:
<Channel list>
<list>
Example:
CURR:ALIM:UPP? (@1)
Queries the upper safety limit for current at channel 1.
Example:
See
Example "Configuring the safety limit"
7.5.3
Remote sense setting
The
VOLTage:SENSe
subsystem contains the command for setting the remote sense
for the instrument.
[SOURce:]VOLTage:SENSe[:SOURce]
.............................................................................122
[SOURce:]VOLTage:SENSe[:SOURce]
<arg0>[, <Channel list>]
[SOURce:]VOLTage:SENSe[:SOURce]?
[<Channel list>]
Sets remote sense detection.
Parameters:
<arg0>
AUTO | EXT | <list>
AUTO
If remote sense detection is set "AUTO", the detection and ena-
bling of the voltage sense relay automatically kicks in when the
connection of remote sense wires (S+, S-) to the input of the
load is applied.
For query return:
AUTOOFF - Remote sense is configured as automatic detection
and currently disabled.
AUTOON - Remote sense is configured as automatic detection
and currently active.
Configuration commands