Instrument functions
R&S
®
NGP800
98
User Manual 5601.5610.02 ─ 09
4. Setup the adjustment with instruments illustrated in
5. Follow the on-screen instructions displayed in the
.
Supply the required voltage to the analog input and key in the measured value
from DMM using the on-screen keypad. See
Figure 6-68: Analog in adjustment procedure
6. Confirm the entry with
.
7. Leave the setup connections as open.
Select
to start the adjustment automatically.
8. If adjustment is successful, the R&S
NGP800 displays a message to indicate that
the adjustment is successful.
The R&S
NGP800 overwrites the factory or the last analog in adjustment.
Figure 6-69: Analog in adjustment successful message
9. If adjustment failed after repeated tries, contact your local service partner for sup-
port.
Figure 6-70: Analog in adjustment failure message
Adjustment