Remote control commands
R&S
®
NGP800
172
User Manual 5601.5610.02 ─ 09
Example:
LOG 0
LOG:FNAM?
-> ""
LOG 1
LOG:FNAM?
-> "/int/logging/log-20190318T1141853.407.csv"
Enables the data logging and queries the data log filename.
LOG:INTerval
<Set new value>[, <Return min or max>]
LOG:INTerval?
[<Return min or max>]
Sets or queries the data logging measurement interval. The measurement interval
describes the time between the recorded measurements.
Setting parameters:
<interval>
<numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Measurement interval in the range of 0.008 s to 600 s.
MIN | MINimum
Minimum measurement interval is set at 0.008 s.
MAX | MAXimum
Maximum measurement interval is set at 600 s.
Default unit: s
Parameters for setting and query:
<Return min or max> MIN | MINimum | MAX | MAXimum
Returns the measurement interval.
Example:
LOG:INT 10
LOG:INT?
-> 10
LOG:LOCation
[<>]
LOG:LOCation?
[<>]
Sets or queries the logging location.
Parameters for setting and query:
<location?
INT | EXT | DEF
INT
Internal location, i.e. "int/location/".
EXT
External location, i.e. "USB1A/lNGP/ocation/".
DEF
Default location, i.e. "int/location/".