Remote control commands
R&S
®
NGP800
116
User Manual 5601.5610.02 ─ 09
LOG
For output mode - output the selected logic level when logging is
enabled.
For input mode - Logging is enabled when the selected logic
level is met.
PLEVel
Output the selected logic level when the power level of the
selected channel is greater or equal to the set power level, i.e.
Plevel >= set value.
<value>
Mode value.
TRIGger:DIRection:DIO<IO>
<arg0>
Sets or queries the specified Digital I/O line to function as Trigger Input/Output.
Suffix:
<IO>
.
1..8
Parameters:
<logic>
OUTPut | INPut
*RST:
OUTPut
Example:
TRIGger:DIRection:DIO2 OUT
TRIGger:LOGic:DIO<IO>
<arg0>
Sets or queries the trigger logic (Active High/Active Low) of the specified Digital I/O
line.
Suffix:
<IO>
.
1..8
Parameters:
<logic>
LOW | HIGH
*RST:
HIGH
TRIGger[:ENABle]:DIO<IO>
<arg0>
Sets or queries the enable state of the specified Digital I/O line.
Suffix:
<IO>
.
1..8
Parameters:
<state>
1
Selected Digital /O line is enabled.
0
Selected Digital /O line is disabled.
*RST:
0
Trigger commands