Remote control commands
R&S
®
NGP800
129
User Manual 5601.5610.02 ─ 09
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
<desired stepsize>[,
<Optional default step query>]
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]?
[<Optional default
step query>]
Sets or queries the incremental step size for the
Setting parameters:
<stepsize>
<numeric value> | DEF | DEFault
<numeric value>
Step value in A.
DEF | DEFault
Default value of stepsize.
Range:
0.0001 to 2.000
Increment: 0.0001
*RST:
0.010
Default unit: A
Parameters for setting and query:
<Optional default step
query>
DEF | DEFault
Queries the default voltage step size.
Example:
INST OUT1
CURR:STEP 0.005
CURR:STEP DEF
VOLT:STEP? DEF
-> 0.1000E+00
Returns the default stepsize for current.
See
Example "Configuring the current output"
7.5.6
Combined setting of voltage and current settings
The
APPLy
subsystem provides a command that enables you to set the current and
voltage of a channel in one step.
The combined voltage and current setting command takes approximately 100 ms, i.e.
longer than the setting of a single value.
APPLy
<arg0>[, <arg1>, <arg2>]
Sets or queries the voltage and current value of the selected channel.
Parameters:
<voltage>
numeric | MIN | MINimum | MAX | MAXimum | DEF | DEFault
numeric
Numeric value for voltage in the range of 0.000 to 64.050.
MIN | MINimum
Min voltage at 0.000 V.
Configuration commands