Remote control commands
R&S
®
NGP800
157
User Manual 5601.5610.02 ─ 09
Parameters:
<data>
voltage1, current1, time1, interpolation mode1, voltage2, cur-
rent2, time2, interpolation mode2, ...
Voltage and current settings depending on the instrument type.
If the interpolation mode is sets to 1, it indicates that the mode is
activated. If the interpolation mode is sets to 0, it indicates that
the mode is not activated.
Example:
INST OUT1
ARB:DATA 10,1,0.5,0
Defines one arbitrary point with: Voltage1 = 10 V and Current1 =
1 A, Time1 = 500 ms and Interpolation mode1 = 0 (disabled).
ARB:DATA?
-> 10.000, 1.000, 0.50, 1
Returns defined arbitrary points for the previous selected chan-
nel.
See
Example "Configuring an arbitrary sequence"
ARBitrary:SEQuence:ENDPoint?
Queries the total number of points of the arbitrary sequence.
Usage:
Query only
ARBitrary:FNAMe
<>[, <>]
ARBitrary:FNAMe?
<>[, <>]
Sets or queries the file name and storage location for the QuickArb function.
Parameters for setting and query:
<filename>
Filename of the QuickArb function.
<location>
INT | EXT | DEF
INT
Internal memory
EXT
USB stick
DEF
Internal memory
Example:
ARB:FNAM "01.CSV"
ARB:FNAM? INT
-> "01.CSV"
ARBitrary:LOAD
Loads an arbitrary table from a file (filename specified with
ARB:FNAM
)
Advanced operating commands