Remote control commands
R&S
®
NGP800
166
User Manual 5601.5610.02 ─ 09
CALibration:CURRent:DATA
<arg0>
Set the DMM reading after setting the output current level in channel adjustment proc-
ess.
Parameters:
<current>
Measured value from DMM.
CALibration:CURRent:IMAX
Sets the output current to high value 100 % of Imax during current adjustment.
Usage:
Event
CALibration:CURRent:IMIN
Sets the output current to low value 1 % of Imax during current adjustment.
Usage:
Event
CALibration:DATE?
Returns the channel adjustment date.
Usage:
Query only
CALibration:END
Ends the channel adjustment.
Usage:
Event
CALibration:FACTory:RESTore
Restores the factory channel adjustment.
Usage:
Event
CALibration:SAVE
Saves the channel adjustment.
Usage:
Event
CALibration:STATe?
Returns the current state of channel adjustment.
Advanced operating commands