Instrument functions
R&S
®
NGP800
63
User Manual 5601.5610.02 ─ 09
Figure 6-18: Digital trigger settings for pin 1
5. Set the required pin "Direction".
Depending on the pin direction, different operating modes are available for trigger
settings.
See
and
.
6. Set the required "Active Level " and "Channel" settings.
7. Press the [Settings] key to return to device menu dialog.
8. Set the "Digital I/O Trigger" to "ON" to enable the digital I/O trigger.
The R&S
NGP800 monitors the digital I/O trigger conditions and feedback to the
respective channels or pins.
6.7
Advanced features
QuickArb function
If QuickArb function of a selected channel is enabled, the respective channel voltage,
current setting and safety limit settings are disabled.
See
Chapter 6.1, "Setting the channels voltage and current"
The QuickArb and EasyRamp are two functions which can be used to control the
waveform of voltage and current output.
6.7.1
QuickArb
The QuickArb function allows you to generate freely programmable waveforms which
can be reproduced within the
1. Press [Settings] key.
Advanced features