Remote control commands
R&S
®
NGP800
161
User Manual 5601.5610.02 ─ 09
1
EasyRamp function is activated.
*RST:
0
Parameters for setting and query:
<Channel list>
<list>
Example:
INST OUT1
VOLT:RAMP ON
VOLT:RAMP?
-> 1
EasyRamp function of Ch1 is activated
Example:
VOLT:RAMP ON, (@1)
VOLT:RAMP? (@1)
Sets and queries the state of ramp function at channel 1.
[SOURce:]VOLTage:RAMP:DURation
<New value for voltage>[, <Channel list>]
[SOURce:]VOLTage:RAMP:DURation?
[<Channel list>]
Sets or queries the duration of the voltage ramp.
Parameters:
<duration>
<numeric value> | MIN | MINimum | MAX | MAXimum | DEF |
DEFault
<numeric value>
Duration of the ramp function in seconds.
MIN | MINimum
Minimum duration of the ramp function at 0.00 s.
MAX | MAXimum
Maximum duration of the ramp function at 60.00 s.
DEF | DEFault
Default duration of the ramp function at 0.01 s.
Range:
0.01 to 60.00
*RST:
0.01
Default unit: s
Parameters for setting and query:
<Channel list>
<list>
Example:
VOLT:RAMP:DUR 4
VOLT:RAMP:DUR?
-> 4
Duration of the ramp function is set at 4 s.
Example:
VOLT:RAMP:DUR 4, (@1)
VOLT:RAMP:DUR? (@1)
Sets and queries the duration of ramp function at channel 1.
Advanced operating commands