Instrument functions
R&S
®
NGP800
64
User Manual 5601.5610.02 ─ 09
The R&S
NGP800 displays the device/channel menu window.
2. Select the required channel tab to configure QuickArb function.
The R&S
NGP800 displays the selected channel menu.
3. Select "Arbitrary" from the menu.
The R&S
NGP800 displays the "Arbitrary" dialog.
Figure 6-19: Arbitrary dialog
4. Select any of the rows to load or change the arbitrary file in the arbitrary sequence.
Up to eight arbitrary files with a maximum of 1024 data points can be loaded.
The R&S
NGP800 opens a dialog to select the source and file location.
5. Select the required source and file location.
6. Select "Select" to load the selected file.
The R&S
NGP800 loads the selected arbitrary file.
Figure 6-20: Select arbitrary file location
7. Select delete and up/down button
to navigate the arbitrary test sequence.
8. Select "Load Sequence" to load the arbitrary test sequence.
9. Set "Seq. Rep." and "End Behavior" to configure the arbitrary sequence behavior.
● "Seq. Rep.": Repetition cycle for the arbitrary sequence
● "End Behavior": End behavior of the automation of arbitrary function
Advanced features