background image

6

Subject to change without notice

I n s t a l l a t i o n   a n d   s a f e t y   i n s t r u c t i o n s          

 1  Installation and safety instructions

1.1.  Setting up the instrument

As can be seen from the fi gures, the handle can be set into 
different positions:

A and B = carrying
C = horizontal operating
D and E = operating at different angles
F = handle removal 
G = operating using the feet‘s, batch use and for shipping in 
original packaging

STOP

 Attention!

 

When changing the handle position, the instrument 
must be placed so that it cannot fall (e.g. placed on 
a table). Then the handle locking knobs must be 
simultaneously pulled outwards and rotated to the 
required position. Without pulling the locking knobs 
they will latch in into the next locking position.

Removal/fi tting of the handle:

 The handle can be removed in 

position F, pulling the side parts outside the housing. Adding 
the handle works vice versa.

1.2. Safety

The instrument fulfi ls the VDE 0411 part 1 regulations for 
electrical measuring, control and laboratory instruments and 
was manufactured and tested accordingly. It left the factory in 
perfect safe condition. Hence it also corresponds to European 
Standard EN 61010-1 and International Standard IEC 1010-1. 
In order to maintain this condition and to ensure safe opera-
tion the user is required to observe the warnings and other 
directions for use in this manual. Housing, chassis as well as 
all measuring terminals are connected to safety ground of the 
mains. All accessible metal parts were tested against the mains 
with 2200 V

DC

. The instrument conforms to safety class  I. The 

oscilloscope may only be operated from mains outlets with a 
safety ground connector. The mains plug has to be installed 
prior to connecting any signals. It is prohibited to separate the 
safety ground connection. If suspected that safe operation may 
not be guaranteed do not use the instrument any more and lock 
it away in a secure place. 

Safe operation may be endangered if any of the following was 
noticed:

–  in case of visible damage. 
–  in case loose parts were noticed 
–  if it does not function any more. 
–  after prolonged storage under unfavourable conditions (e.g. 

like in the open or in moist atmosphere). 

–  after any improper transport (e.g. insuffi cient packing not 

conforming to the minimum standards of post, rail or trans-
port fi rm) 

1.3. Correct 

operation

Please note: This instrument is only destined for use by person-
nel well instructed and familiar with the dangers of electrical 
measurements. For safety reasons the oscilloscope may only 
be operated from mains outlets with safety ground connector. 
It is prohibited to separate the safety ground connection. The 
plug must be inserted prior to connecting any signals. 

The oscilloscope is destined for operation in industrial, business, 
manufacturing, and domestic sites.

1.4. Ambient 

conditions

Operating ambient temperature: +5 °C to +40 °C. During trans-
port or storage the temperature may be –20 °C to +70°C. Please 
note that after exposure to such temperatures or in case of 
condensation, proper time must be allowed until the instrument 
has reached the permissible temperature, and until the con-
densation has evaporated before it may be turned on! Ordinarily 
this will be the case after 2 hours. The oscilloscope is destined 
for use in clean and dry environments. Do not operate in dusty 
or chemically aggressive atmosphere or if there is danger of 
explosion. The any operating position may be used, however, 
suffi cient ventilation must be ensured. Prolonged operation 
requires the horizontal or inclined position.

STOP

  Do not obstruct the ventilation holes! 

Specifi cations are valid after a 30 minute warm-up period at 23 
degr. C (tolerance ±2 degr. C). Specifi cations without tolerances 
are average values.

1.5. Warranty 

and 

repair

HAMEG instruments are subjected to a strict quality control. 
Prior to leaving the factory, each instrument is burnt in for 10 
hours. By intermittent operation during this period almost all 
defects are detected. Following the burn in, each instrument is 
tested for function and quality, the specifi cations are checked 

Carrying positions

Operating positions

Removing the handle (Pos. F)

Stacking positions

E

D

C

A

B

A

B

C

D

E

F

G

Positions of the instrument

G

Summary of Contents for Hameg HMO3522

Page 1: ...350 MHz Digital Oscilloscope with FFT and mixed signal option HMO3522 3524 Manual English...

Page 2: ...different For measurementoperationthefollowinghintsandconditionsregardingemission and immunity should be observed 1 Data cables For the connection between instrument interfaces and external devices c...

Page 3: ...1 Coupling 15 4 2 Sensitivity Y Positioning and Offset 16 4 3 Bandwidth Limit and Signal Inversion 16 4 4 Probe attenuation selection 16 5 Horizontal System Time Base 17 5 1 Capturing modes RUN and ST...

Page 4: ...mV 5V div into 1M 50 Offset control 0 2 20V 12div x axis display range Trigger modes slope video pulsewidth logic delayed event FFT for spectral analysis 6 digit counter Autoset automeasurement formul...

Page 5: ...tenuation ID HZ350 Dual Interface USB RS 232 HO720 CD Optional accessories HO730 Dual Interface Ethernet USB HO740 Interface IEEE 488 GPIB galvanically isolated HZ46 4RU 19 Rackmount Kit HZ355 Slimlin...

Page 6: ...f visible damage in case loose parts were noticed if it does not function any more after prolonged storage under unfavourable conditions e g like in the open or in moist atmosphere after any improper...

Page 7: ...ring circuits are considered not connected to the mains if a suitable isolation transformer ful lling safety class II is used Measurements on the mains are also possible if suitable probes like curren...

Page 8: ...interface STOP Please note the connectors for the active logic probes HO3508 51 52 are solely for theses probes Connecting anything else could destroy the inputs 2 2 Control panel The controls on the...

Page 9: ...nel actually being displayed these may be selected using the soft keys 3 Below the graticule measurement results of parameters and cursors the settings of the activated vertical channels of the refere...

Page 10: ...e the rst one will follow 2 7 Basic setting and integrated help Basic settings like language for user interface and help mis cellaneous settings and interface settings can be set using the menu which...

Page 11: ...r the date and build information Fig 2 11 Updating menu and information window Now choose which to update the rmware or the help function If both are to be updated it is recommended to rst update the...

Page 12: ...t ground connection possible to the PROBE ADJUST output Connect the ground cable to the probe and the tip to the top of the probe Now connect the compensation box BNC connector to the input connector...

Page 13: ...ner of the screen In this case the V cursor has se lected the voltages at the two cursor positions their difference and the time difference between the positions will be shown The cursors will be swit...

Page 14: ...ne or two prede ned mathematical functions A quick setting of mathematical functions is possible by selecting the menu item at the bottom This mode allows you to select the addition or subtraction of...

Page 15: ...4 1 Front panel area with vertical system controls SCALE VOLTS DIV Y POSITION CHANNEL CONTROL FINE 62 K304 1 1 5 0 1 0 MENU CH1 CH2 CH3 POD1 MATH REF XY CH4 POD2 18 B 19 20 22 23 24 21 25 26 27 By pu...

Page 16: ...ob The offset voltage will be added to the signal at the vertical ampli er input offsetting it by that amount from the zero position The possible amount of offset depends on the Volts div setting chos...

Page 17: ...envelope of the signal will be displayed Average Functions only with repetitive signals The universal knob in the CURSOR MENU section of the front panel is used to set the number of signal periods for...

Page 18: ...tly used functions TYPE selects the type of trigger SLOPE PULSE LOGIC VIDEO and the B TRIGGER SLOPE selects the slope polarity SOURCE opens the menu for the selection of the trigger source FILTER open...

Page 19: ...he trigger such that rst condition A must be met and then another condition B before the trigger will respond refer to Fig 6 3 E g it is possible to de ne a source channel and a level of 120mV on the...

Page 20: ...me available which can be shifted up and down with the universal knob A detailed description will follow in the next chapter There are 3 more menu items on the rst page DOTS ONLY The respective soft m...

Page 21: ...ion to the time elapsed This mode can be selected in the soft menu which will open upon pushing the key INTENS PERSIST the signal intensity can be changed also in this menu Fig 7 2 Menu for setting th...

Page 22: ...will terminate the XY function if it was active If no or another menu should be active it will be necessary to push the XY key twice for terminating the function 8 Measurements There are two different...

Page 23: ...SORMODEopensthemenuitemGLUE TO Thismode can be turned on or off If activated the cursors will glue to the signal i e they will automatically follow all changes of the position and scaling controls and...

Page 24: ...mode offers the following 5 parameters which are directly displayed in the signal Maximum voltage Mean voltage Minimum voltage Rise time Fall time 4 additional parameters will be displayed in the righ...

Page 25: ...the associated menu In Fig 9 4 in formula MA1 a current of 100 A is added to the channel 1 In the menu for entering constants it is possible to choose from the following list of constants executed by...

Page 26: ...beshown top left the information about the Zoom and position between both grids and the information about the FFT display span and center frequency is shown below the larger area One of these displays...

Page 27: ...menu key next to STORE In order to recall stored instrument settings call the main instrument settings menu and select LOAD by pushing the respective soft menu key The data manager will open use the m...

Page 28: ...addition to references the pure A D converter data can be stored however only on external USB sticks not internally The following formats are available Binary format A binary data set may contain byte...

Page 29: ...item is called FORMULARIES selection of this menu item will call a submenu which allows you to move sets of for mulas between the internal memory and the USB stick as well as to import or export such...

Page 30: ...menu eld below by the respective soft key These 6 criteria are available ti t The duration of the bit pattern which will generate the trigger is unequal to the reference time ti t The duration of the...

Page 31: ...his method all channels may be individually positioned and sized If POD was activated and if the MENU key in the Channel Control section of the front panel was pushed the menu for setting the trigger...

Page 32: ...e 15 Fig 4 1 Front panel area with vertical system controls 15 Fig 4 2 Short menu for the vertical settings 15 Fig 4 3 Vertical offset in the extended menu 16 Fig 5 1 Control panel of the horizontal s...

Page 33: ...33 Subject to change without notice A p p e n d i x...

Page 34: ...34 Subject to change without notice Appendix...

Page 35: ...35 Subject to change without notice A p p e n d i x...

Page 36: ...tra e 6 HAMEG Instruments GmbH D 63533 Mainhausen A Rohde Schwarz Company Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg com Oscilloscop...

Reviews: