Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-43
16.
:FETCh:HARMonics:FREQuency:ALL?
Syntax
:FETCh:HARMonics:FREQuency:ALL?
Function
The query returns the frequency of 10 harmonics at comma
intervals in Hz. The first harmonic is the fundamental wave.
Explanations
The command returns “0” for those unmeasured harmonics if
less than 10 harmonics to measrure.
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value Such as 45500000,91000000,136500000,182000000, 227500000,
273000000,318500000,364000000,409500000,455000000
17.
:FETCh:HARMonics:FREQuency? <n>
Syntax
:FETCh:HARMonics:FREQuency? <n>
Function
Queries the frequency of specified harmonic in Hz, such
as :FETC:HARM:FREQ? 1
Explanations
<n> is an integer wthin 1 and 10.
The query returns “---” if the harmonic to be read has no data.
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value Such as 45500000
18.
:FETCh:HARMonics:FUNDamental?
Syntax
:FETCh:HARMonics:FUNDamental?
Function
The query returns the fundamental wave freuqncy in Hz.
Explanations
The commands is the same
as :FETCh:HARMonics:FREQuency? 1
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value Such as 45500000
19.
:FETCh:OBWidth?
Syntax
:FETCh:OBWidth?
Function
The query returns 2 values at comma intervals: Occupied
BandWidth (Hz) and Transmit Freq Error (Hz).
Explanations The command transmits the data from the latest measurement to
the ouput buffer.
Return Value Such as 1860000,40000
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...