Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-23
Syntax
:CALCulate:MARKer<n>:X:SPAN <param>
:CALCulate:MARKer<n>:X:SPAN?
Function
Sets the X value of the marker span on the related trace in “Span
Pair” marker mode. The unit of X value corresponds to the unit of
the current marker trace.
Such as :CALC:MARK1:X:SPAN 500000000
The query returns “500000000”.
Explanations
<n> is 1, 2, 3 or 4.
<param> is within the range of X.
If current readout mode is “Frequency”, the range cannot
exceed the sweep frequency and unit input of Hz (default),
kHz, MHz or GHz is available.
If current readout mode is “
△
time”, the range cannot exceed
the sweep time and unit input of s (default), us, ms or ks is
available.
Front Panel
Marker, Span Pair, Span
52.
:CALCulate:MARKer<n>:X:STARt
Syntax
:CALCulate:MARKer<n>:X:STARt <param>
:CALCulate:MARKer<n>:X:STARt?
Function
Sets the X value of the reference marker on related trace in “Delta
Pair” marker mode. The unit of X value corresponds to the unit of
the current marker trace.
Such as :CALC:MARK1:X:STARt 1320000000
The query returns “1320000000”.
Explanations
<n> is 1, 2, 3 or 4.
<param> is within the range of X.
If current readout mode is “Frequency”, the range cannot
exceed the sweep frequency and unit input of Hz (default),
kHz, MHz or GHz is available.
If current readout mode is “
△
time”, the range cannot exceed
the sweep time and unit input of s (default), us, ms or ks is
available.
Front Panel
Marker, Delta Pair, Ref
53.
:CALCulate:MARKer<n>:X:STOP
Syntax
:CALCulate:MARKer<n>:X:STOP <param>
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...