Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-93
Front Panel
Meas, OBW
Meas Setup, Avg Mode, Exp/Repeat
76.
[:SENSe]:OBWidth:MAXHold:STATe
Syntax
[:SENSe]:OBWidth:MAXHold:STATe OFF|ON|0|1
[:SENSe]:OBWidth:MAXHold:STATe?
Function
Turns on or off the Max Hold function of the Occupied Bandwidth
measurement.
The query returns “0” or “1”.
Default
OFF|0
Front Panel
Meas, OBW
Meas Setup, Max Hold, On/Off
77.
[:SENSe]:OBWidth:FREQuency:SPAN
Syntax
[:SENSe]:OBWidth:FREQuency:SPAN <freq>
[:SENSe]:OBWidth:FREQuency:SPAN?
Function
Sets the span of the Occupied Bandwidth measurement, the default
unit is Hz, such as
:OBW:FREQ:SPAN 1000000 or :OBW:FREQ:SPAN 1MHZ
The query returns “1000000”.
Explanations
<freq> ranges from 100 Hz to 3 GHz.
Default
2MHz
Front Panel
Meas, OBW
Meas Setup, Span
78.
[:SENSe]:OBWidth:PERCent
Syntax
[:SENSe]:OBWidth:PERCent <real>
[:SENSe]:OBWidth:PERCent?
Function
Sets the percent of the signal power in the whole frequency span
power, such as
:OBW:PERC 90
The query returns “9.01”.
Explanations
<real> ranges from 1 to 99.99.
Default
99 (99%)
Front Panel
Meas, OBW
Meas Setup, Power Ratio
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...