Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-33
1.
:DISPlay:AFUnction:POSition
Syntax
:DISPlay:AFUnction:POSition BOTTom|CENTer|TOP
:DISPlay:AFUnction:POSition?
Function
Sets the position of active function area.
The query returns “BOTT”, “CENT” or “TOP”.
Explanations This setting is persistent even throuth a power cycle.
Default
TOP
Front Panel
Display, Active Fctn, Top/Center/Bottom
2.
:DISPlay:ANNotation:CLOCk
Syntax
:DISPlay:ANNotation:CLOCk[:STATe] OFF|ON|0|1
:DISPlay:ANNotation:CLOCk[:STATe]?
Function
Sets the clock status.
The query returns “0” or “1”.
Explanations This setting is persistent even throuth a power cycle.
Default
ON|1
Front Panel
System, Time/Date, Time/Date, On/Off
3.
:DISPlay:BRIGtness
Syntax
:DISPlay:BRIGtness <integer>
:DISPlay:BRIGtness?
Function
Sets the screen brightness, such as
:DISP:BRIG 6
The query returns “6”.
Explanations <integer> ranges from 0 to 7.
Default
6
Front Panel
Display, Brightness
4.
:DISPlay:ENABle
Syntax
:DISPlay:ENABle OFF|ON|0|1
:DISPlay:ENABle?
Function
Enables or disables the screen display. The refresh will be stopped
when screen display is disabled, pressing ESC to unlock.
The query returns “0” or “1”.
Default
ON|1
Front Panel
Display, Scr State, On/Off
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...