Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-13
The command will refers all activated markers on condition that
the frequency counter is enabled.
Default
OFF|0
Front Panel
Marker Fctn, Freq Count, State, On/Off
20.
:CALCulate:MARKer:FCOunt:X?
Syntax
:CALCulate:MARKer:FCOunt:X?
Function
The query returns the readout of the frequency counter, such as
1500335500, (in Hz).
Explanations The query returns “9000000000000000 (9e15)” when the
frequency counter is disabled.
21.
:CALCulate:MARKer<n>:FUNCtion
Syntax
:CALCulate:MARKer<n>:FUNCtion NDB|NOISe|OFF
:CALCulate:MARKer<n>:FUNCtion?
Function
Selects a measurement type for the specified marker.
The query returns “NDB”, “NOIS” or “OFF”.
Explanations
<n> is 1, 2, 3 or 4.
NDB: N dB BW measurement.
NOISe: Noise measurement.
OFF: turns off all measurements.
Default
OFF|0
Front Panel
Marker Fctn, Noise Mkr/N dB BW/
Function Off
22.
:CALCulate:MARKer<n>:MAXimum:MAX
Syntax
:CALCulate:MARKer<n>:MAXimum:MAX
Function
Executes a peak search according to the mode selected
by :CALCulate:MARKer<n>:PEAK:SEARch:MODE.
Explanations <n> is 1, 2, 3 or 4.
Front Panel
Peak, Search Para, Peak Search, Max
23.
:CALCulate:MARKer<n>:MAXimum:LEFT
Syntax
:CALCulate:MARKer<n>:MAXimum:LEFT
Function
Searches the nearest peak located to the left side of the current
peak and meets the searching condition. The peak is then identified
with a marker.
Explanations
<n> is 1, 2, 3 or 4.
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...