Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-95
:TOI:AVER:COUN 100
The query returns “100”.
Explanations
<integer> ranges from 1 to 1000.
Default
10
Front Panel
Meas, TOI
Meas Setup, Avg Num
86.
[:SENSe]:TOI:AVERage[:STATe]
Syntax
[:SENSe]:TOI:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:TOI:AVERage[:STATe]?
Function
Turns on or off the average function of the TOI measurement.
The query returns “0” or “1”.
Default
OFF|0
Front Panel
Meas, TOI
Meas Setup, Avg Num, On/Off
87.
[:SENSe]:TOI:AVERage:TCONtrol
Syntax
[:SENSe]:TOI:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:TOI:AVERage:TCONtrol?
Function
Selects an average mode of the TOI measurement from EXP or REP.
The query returns “EXP” or “REP”.
Explanations
In Exp mode, the result is from the exponent average calculation
of the current N (specified by [:SENSe]:TOI:AVERage:COUNt)
measurements.
In REP mode, the result is the arithmetic average of the current N
(specified by [:SENSe]:TOI:AVERage:COUNt) measurements.
Default
EXPonential
Front Panel
Meas, TOI
Meas Setup, Avg Num, Exp/Repeat
88.
[:SENSe]:TOI:FREQuency:SPAN
Syntax
[:SENSe]:TOI:FREQuency:SPAN <freq>
[:SENSe]:TOI:FREQuency:SPAN?
Function
Sets the span of the TOI measurement, the default unit is Hz, such as
:TOI:FREQ:SPAN 1000000 or :TOI:FREQ:SPAN 1MHZ
The query returns 1000000.
Explanations
<freq> ranges from 100 Hz to 3 GHz.
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...