:CALCulate:BANDwidth:NDB
Syntax
:CALCulate:BANDwidth:NDB <rel_ampl>
:CALCulate:BANDwidth:NDB?
Function
Sets the N value of the N dB BW measurement, such as
:CALC:BAND:NDB -4
The query returns “-4.00”.
Explanations
<rel_ampl> ranges from -100 dB to 100 dB.
Default
-3 dB
Front Panel
Marker Fctn, N dB BW
2.
:CALCulate:BANDwidth:RESult?
Syntax
:CALCulate:BANDwidth:RESult?
Function
The query returns the measured bandwidth (in Hz) according to the
N value specified by command
:CALCulate:BANDwidth:NDB, such as 80000000.
Explanations The query returns “---” if the returned value is unavailable.
Front Panel
Marker Fctn, N dB BW
3.
:CALCulate:LLINe:ALL:DELete
Syntax
:CALCulate:LLINe:ALL:DELete
Function
Deletes the current edited limit line.
Front Panel
Meas, Pass/Fail
Meas Setup, Limit, Upper, Del Limit
Meas Setup, Limit, Lower, Del Limit
4.
:CALCulate:LLINe:CONTrol:DOMain
Syntax
:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME
:CALCulate:LLINe:CONTrol:DOMain?
Function
Specifies the unit of Frequency or Time as the unit of X-axis.
The query returns “FREQ” or “TIME”
.
Explanations
The command works for both upper and lower limit.
Changes in unit of X-axis may delete all edited points under the
current limit line.
Default
FREQuency
Front Panel
Meas, Pass/Fail
Meas Setup, X-axis, Freq/Time
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...