Remote Commands
R&S
®
ZNH
242
User Manual 1334.5985.02 ─ 02
Example:
DISP:TRAC:Y:ADJ
Adjusts the y-axis.
Usage:
Event
DISPlay<1...2>:IMPedance:REFerence:POSition
<RefPosition>
This command sets the reference impedance for the smith chart measurement format.
The impedance can be between 1mΩ and 10 kΩ.
The numeric suffix at DISPlay selects the measurement screen in dual trace mode and
is in the range <1...2>.
Suffix:
<1...2>
.
1...2
Parameters:
<RefPosition>
Numeric value in the rnage from 1 mΩ to 10 kΩ.
*RST:
50
Default unit: OHM
Example:
DISP:IMP:REF:POS 75 OHM
Sets the reference impedance to 75 Ohm.
DISPlay<1...2>:LOSS:REFerence
<RefLevel>
This command defines the reference level for the cable loss measurement format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<RefLevel>
Numeric value in the range from -100 dB to 100 dB.
*RST:
0 dB
Default unit: dB
Example:
DISP:LOSS:REF 10
Defines a reference level of 10 dB.
DISPlay<1...2>:LOSS:REFerence:POSition
<RefPosition>
This command defines the position of the reference value in the diagram for the cable
loss measurement format.
Each step shifts the reference position one grid line up or down.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Remote Control - Commands