Cable and Antenna Test Mode
R&S
®
ZNH
111
User Manual 1334.5985.02 ─ 02
Figure 7-4: RL + Trans measurement display
You can perform reflection measurements on the complete system or on individual
components of the system. If you measure while two or more system components are
connected, the results of the reflection measurement are an aggregate over these
components. Therefore you can only see the aggregated magnitude of the reflected
power in a specified frequency range.
To draw conclusions about which component is affected and to determine the location
of the fault, you need to perform further analysis using other measurements.
7.1.1.1
Managing S-Parameters File
The R&S
ZNH is able to export the reflection measurements trace as S1P and S2P file
in Touchstone format 1.1.
To export the file, do the following:
1. Press [SAVE/RECALL] key.
2. Select the "Export" softkey.
The R&S
ZNH opens the "Export" dialog box.