Remote Commands
R&S
®
ZNH
273
User Manual 1334.5985.02 ─ 02
F2PShortnorm
Refl Norm Short calibration routine on port 1 and port 2.
TOSMcal
A full two port calibration routine performs on both test ports.
UOSMcal
Same calibration routine as "TOSMcal" but with unknown
Through standard.
Example:
CAL:STAR? S22Opennorm
Selects and initiates an S22Opennorm calibration and would
return e.g.:
Calibration done
Calibration has been finished. No further actions required.
Usage:
Query only
CALibration<1...2>:STATus?
This command queries if the R&S
ZNH is fully calibrated for the current measurement.
The numeric suffix at CALibration is irrelevant for this command.
This command is a query and therefore has no *RST value.
Suffix:
<1...2>
.
1...2
Parameters:
<Return values>
NORMalized | APPRoximate
NORMalized
Full factory or user calibration
APPRoximate
Approximate calibration: measurement uncertainty must be
anticipated
Example:
CAL:STAT?
Queries the calibration status of the R&S
ZNH.
Usage:
Query only
Working with a DTF List
The following commands configure the DTF peak list. The peak list is available for DTF
measurements only.
List of commands
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DTF:PEAK:RESult?
.......................................274
CALCulate<1...2>:TRANsform:DTFault:PEAK:COUNt?
......................................................274
CALCulate<1...2>:TRANsform:DTFault:PEAK:DATA<1...10>?
............................................ 274
CALCulate<1...2>:TRANsform:DTFault:PEAK[:STATe]
.......................................................275
CALCulate<1...2>:TRANsform:DTFault:PEAK:THReshold
..................................................275
CALCulate<1...2>:MARKer<1...6>:FUNCtion:DTF:PEAK:THReshold
.................................. 275