Remote Commands
R&S
®
ZNH
214
User Manual 1334.5985.02 ─ 02
Note that you have to set a logarithmic scaling before you can use this command with
DISPlay<1...2>:MAGNitude:Y:SPACing
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<DisplayRange>
Numeric value of the display range.
The number you enter is rounded up to the next possible display
range. For example, if you enter 9, the R&S
ZNH automatically
sets the display range to 10.
*RST:
100 dB
Default unit: dB
Example:
DISP:IMAG:Y:SCAL 50 DB
Defines a display range of 50 dB.
DISPlay<1...2>:MLINear:REFerence
<RefLevel>
This command defines the reference value for the linear scale of the magnitude mea-
surement format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<RefLevel>
Numeric value of the reference level.
*RST:
0 dB
Default unit: dB
Example:
DISP:MLIN:REF -10
Defines a reference level of -10 dB.
DISPlay<1...2>:MLINear:REFerence:POSition
<RefPosition>
This command defines the position of the reference value in the diagram for the linear
part of the magnitude measurement format.
Each step shifts the reference position one grid line up or down.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<RefPosition>
Numeric value of the reference position.
*RST:
10
Example:
DISP:MLIN:REF:POS 5
Moves the reference to the fifth grid line from the bottom.
Remote Control - Commands