Remote Control
R&S
®
FSPN
338
User Manual 1179.4363.02 ─ 01
Manual operation:
See
INPut<ip>:SELect
<Source>
This command selects the signal source for measurements, i.e. it defines which con-
nector is used to input data to the R&S
FSPN.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Source>
RF
Radio Frequency ("RF INPUT" connector)
*RST:
RF
Manual operation:
See
INPut<1|2>:LOSCillator:SOURce
<Location>
This command selects the type of local oscillator in the test setup.
Suffix:
<1|2>
.
1|2
irrelevant
Parameters:
<Location>
EXTernal
External local oscillator connected to the "LO AUX Input" of the
R&S
FSPN.
INTernal
Internal local oscillator of the R&S
FSPN.
*RST:
INTernal
Example:
//Select external oscillator
CONF:PNO:MEAS RES
INP:LOSC:SOUR EXT
Manual operation:
See
INPut<1|2>:LOSCillator:SOURce:EXTernal:LEVel
<Level>
This command selects the level of an external LO signal that is fed into the R&S
FSPN.
Suffix:
<1|2>
.
1|2
irrelevant
Parameters:
<Level>
HIGH
LO signal with high level characteristics.
Common Measurement Settings