Remote Control
R&S
®
FSPN
325
User Manual 1179.4363.02 ─ 01
●
To select the Spot Noise vs Tune measurement, use
●
To select the Transient Analysis measurement, use
●
To select other measurements, use
.
Parameters:
<State>
ON | 1
Selects the VCO Characterization measurement.
OFF | 0
Selects the Phase Noise measurement (
CONF:PNO:MEAS
PNO
).
*RST:
OFF
Example:
//Select the phase noise measurement
CONF:VCO:MEAS OFF
Manual operation:
See
11.5.2
Performing Measurements
................................................................................................326
ABORt
This command aborts the measurement in the current channel and resets the trigger
system.
To prevent overlapping execution of the subsequent command before the measure-
ment has been aborted successfully, use the
*OPC?
or
*WAI
command after
ABOR
and
before the next command.
Note on blocked remote control programs
:
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish and the remote
channel to the R&S
FSPN is blocked for further commands. In this case, you must
interrupt processing on the remote channel first in order to abort the measurement.
To do so, send a "Device Clear" command from the control instrument to the
R&S
FSPN on a parallel channel to clear all currently active remote channels. Depend-
ing on the used interface and protocol, send the following commands:
●
Visa
:
viClear()
Now you can send the
ABORt
command on the remote channel performing the mea-
surement.
Measurements and Result Displays