Example:
//Sweep in forward direction
SWE:FORW ON
Manual operation:
See
"Selecting the sweep direction"
[SENSe:]SWEep:FSEGment
This command stops a measurement in any given half decade and resumes the mea-
surement of the next half decade.
Usage:
Event
Manual operation:
See
"Defining cross-correlation parameters"
[SENSe:]SWEep:MODE
<Mode>
This command selects the configuration mode of the half decade table.
Parameters:
<Mode>
MANual
Manual mode: allows you to select a custom resolution band-
width and number of cross-correlations for each half decade.
•
Define the RBW for a half decade with
•
Define the number of cross-correlations for a half decade with
NORMal
Automatic mode: the application selects the resolution band-
width and number of cross-correlations based on the RBW and
XCORR factors.
•
Define the RBW factor with
•
Define the XCORR factor with
*RST:
NORMal
Example:
//Automatically select RBW and XCORR based on an RBW fac-
tor of 5 % and an XCORR factor of 1.5.
SWE:MODE NORM
LIST:BWID:RAT 5
LIST:RANG5:XCO 1.5
Manual operation:
See
"Defining the resolution bandwidth"
See
"Defining cross-correlation parameters"
See
"Working with the half decade configuration table"
[SENSe:]SWEep:TIME
<Time>
This command queries the estimated measurement time.
Common Measurement Settings