Remote Control
R&S
®
FSPN
361
User Manual 1179.4363.02 ─ 01
11.6.6.1
Noise Configuration
Remote commands to configure the measurement range described elsewhere:
●
●
.......................................................................................... 361
[SENSe:]LIST:BWIDth[:RESolution]:RATio
........................................................................361
[SENSe:]LIST:RANGe<r>:BWIDth[:RESolution]
.................................................................362
[SENSe:]LIST:RANGe<r>:FREQuency:STARt?
................................................................. 362
[SENSe:]LIST:RANGe<r>:FREQuency:STOP?
..................................................................363
................................................................................... 363
[SENSe:]LIST:BWIDth[:RESolution]:USMall
...................................................................... 364
.................................................................................................364
.............................................................................................364
........................................................................................... 365
...............................................................................................366
[SENSe:]SWEep:XOPTimize[:STATe]
...............................................................................366
[SENSe:]SWEep:XOPTimize:THReshold
..........................................................................366
DISPlay:XGINdicator[:STATe]
<State>
This command turns the cross-correlation gain indicator (gray area in the Noise Spec-
trum result display) on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
ON
Example:
//Remove the gray area from the diagram area.
DISP:XGIN OFF
Manual operation:
See
"Turning the display of the cross-correlation gain indicator
[SENSe:]LIST:BWIDth[:RESolution]:RATio
<Percentage>
This command defines the factor that the resolution bandwidth applied to each half
decade is based on.
Parameters:
<Percentage>
<numeric value>
Range:
0.1 to 30
*RST:
10
Default unit: PCT
Common Measurement Settings