Remote Control
R&S
®
FSPN
366
User Manual 1179.4363.02 ─ 01
Parameters:
<Time>
<numeric value>
Default unit: s
Example:
//Query measurement time
SWE:TIME?
would return, e.g.
0.050432
Example:
//Define measurement time
SWE:TIME 2S
Manual operation:
See
"Estimating the measurement time"
See
"Defining the measurement time"
[SENSe:]SWEep:XFACtor
<Factor>
This command defines the cross-correlation factor (XCORR factor) that defines the
cross-correlation operations in each half decade.
Parameters:
<Factor>
<numeric value>
Range:
1
Example:
//Define an XCORR factor of 10
SWE:XFAC 10
Manual operation:
See
"Defining cross-correlation parameters"
[SENSe:]SWEep:XOPTimize[:STATe]
<State>
This command turns optimization of the cross-correlation operations on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
ON
Example:
SWE:XOPT OFF
Turns off optimization (always performs the full number of opera-
tions).
Manual operation:
See
"Defining cross-correlation parameters"
[SENSe:]SWEep:XOPTimize:THReshold
<Threshold>
This command defines the threshold for the cross-correlation optimization feature.
When the threshold (distance between measured values and gain indicator) is reached
in any given half decade, the measurement in that half decade stops and resumes to
measure the next one.