Common Measurement Settings
R&S
®
FSPN
141
User Manual 1179.4363.02 ─ 01
The remote commands required to configure the frequency deviation calculation are
described in
Chapter 11.6.9.7, "Frequency Deviation"
6.9.8
Trace Configuration
Access
: "Overview" > "Analysis" > "Traces"
You can display several traces in the transient analysis. Most trace functionality is the
same as in the phase noise application. Trace settings specific to the transient analysis
are described below, for more information about trace settings also available in other
measurements, refer to the following topics.
The remote commands required to configure traces are described in
Settings in the "Traces" dialog box described elsewhere:
●
" Trace 1 / Trace 2 / Trace 3 / Trace 4 / Trace 5 / Trace 6 "
●
●
Trace export:
Chapter 7.3.3, "Trace Export and Import"
●
Trace copy:
Chapter 7.3.4, "Copying Traces"
●
Trace mathematics:
Note that trace mathematics are only available in the
diagram.
●
●
The remote commands required to configure traces are described in
6.9.9
Marker Configuration
Access
: "Overview" > "Analysis" > "Markers"
Marker functionality is the same as in other phase noise measurements. For more
information, see
PN Transient Analysis Configuration