Remote Control
R&S
®
FSPN
362
User Manual 1179.4363.02 ─ 01
Example:
//Define an RBW factor of 3 %
LIST:BWID:RAT 3
Manual operation:
See
"Defining the resolution bandwidth"
[SENSe:]LIST:RANGe<r>:BWIDth[:RESolution]
<BandWidth>
This command defines the resolution bandwidth applied in a half decade.
The application of the command depends on the half decade configuration mode
(
):
●
Automatic mode: The RBW can only be queried, because it is calculated based on
an RBW factor.
●
Manual mode: The RBW can be defined directly as an absolute value.
Suffix:
<r>
.
1..n
Parameters:
<BandWidth>
<numeric value>
*RST:
Automatic configuration mode is on.
Default unit: Hz
Example:
//Define an RBW of 1 kHz for the fifth half decade
SWE:MODE MAN
LIST:RANG5:BWID 1KHZ
Manual operation:
See
"Defining the resolution bandwidth"
See
"Working with the half decade configuration table"
[SENSe:]LIST:RANGe<r>:FREQuency:STARt?
This command queries the start frequency offset of a half decade.
Suffix:
<r>
.
1..n
Return values:
<Frequency>
<numeric value>
Default unit: Hz
Example:
//Query start frequency for fifth half decade
LIST:RANG5:FREQ:STAR?
would return, e.g.
100000
Usage:
Query only
Manual operation:
See
"Working with the half decade configuration table"
Common Measurement Settings